A novel approach for Venezia ovoid commissioning with a comprehensive analysis of source positions in high-dose-rate brachytherapy

Published:October 18, 2022DOI:https://doi.org/10.1016/j.brachy.2022.08.017

      Abstract

      Purpose

      The lunar design of a Venezia ovoid makes commissioning of the applicator very challenging with traditional autoradiography. In this study, we propose a novel solution to ovoid commissioning and a quality assurance (QA) workflow to effectively assess the entire source path.

      Methods and materials

      A two-step commissioning process, using electron radiation and radiochromic films, was developed to verify the most distal source position. The ovoid was first attached to a film and was irradiated with a 12 MeV linac beam. This process was repeated on a separate, unexposed film, followed by irradiating it with a HDR source at the most distal position. Two lengths, including the ovoid thickness and the distance between the irradiated spot and the ovoid's outer surface, were obtained from the films’ intensity maps. The offset value was calculated from the subtraction of the two measured lengths. Besides acquiring the offset, a source positional simulator (SPS) and a series of planar x-rays from two orthogonal orientations were used to characterize source movement within the ovoid.

      Results

      Compared to x-ray-based autoradiography, the electron exposure significantly improved the ovoid's visibility on film. Our approach did not use surrogate, which further improved measurement outcomes by decreasing inherent uncertainties. The SPS results suggested the source movement was complex within the cervicovaginal area, but it was predictable with the proposed QA workflow.

      Conclusion

      We introduced a novel, surrogate-free method to commission the Venezia ovoid, which facilitated a manual applicator reconstruction. Additionally, we recommended QA multiple source positions to safely use the ovoid in clinical settings.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Brachytherapy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Franziska Walter
        • Maihöfer Cornelius
        • Schüttrumpf Lars
        • et al.
        Combined intracavitary and interstitial brachytherapy of cervical cancer using the novel hybrid applicator Venezia: Clinical feasibility and initial results.
        Brachytherapy. 2018; 17: 775-781
        • Manon Kissel
        • Fournier-Bidoz Nathalie
        • Henry Olivier
        • et al.
        Venezia applicator with oblique needles improves clinical target volume coverage in distal parametrial tumor residue compared to parallel needles only.
        Journal of Contemporary Brachytherapy. 2021; 13: 24-31
        • Monica Serban
        • Kirisits Christian
        • de Leeuw Astrid
        • et al.
        Ring versus ovoids and intracavitary versus intracavitary-interstitial applicators in cervical cancer brachytherapy: results from the EMBRACE I study.
        International Journal of Radiation Oncology Biology Physics. 2020; 106: 1052-1062
        • Erich Schnell
        • Thompson Spencer
        • Ahmad Salahuddin
        • et al.
        Dosimetric considerations when utilizing Venezia, Capri, Rotte double tandem, and tandem and ring with interstitial needles for the treatment of gynecological cancers with high dose rate brachytherapy.
        Medical Dosimetry. 2020; 45: 21-27
        • N Nomden Christel
        • Leeuw Astrid AC de
        • Moerland Marinus A
        • et al.
        Clinical use of the Utrecht applicator for combined intracavitary/interstitial brachytherapy treatment in locally advanced cervical cancer.
        International Journal of Radiation Oncology Biology Physics. 2012; 82: 1424-1430
        • Keller A
        • L Rodríguez-López J
        • Patel A K
        • et al.
        Early outcomes after definitive chemoradiation therapy with Vienna/Venezia hybrid high-dose rate brachytherapy applicators for cervical cancer: A single-institution experience.
        Brachytherapy. 2021; 20: 104-111
        • Harkenrider Matthew M
        • Surucu Murat
        • Harmon Grant
        • et al.
        Early outcomes and impact of a hybrid IC/IS applicator for a new MRI-based cervical brachytherapy program.
        Brachytherapy. 2018; 17: 187-193
        • Lars Fokdal
        • Sturdza Alina
        • Mazeron Renaud
        • et al.
        Image guided adaptive brachytherapy with combined intracavitary and interstitial technique improves the therapeutic ratio in locally advanced cervical cancer: Analysis from the retroEMBRACE study.
        Radiotherapy and Oncology. 2016; 120: 434-440
        • Ravinder Nath
        • Anderson Lowell L
        • Meli Jerome A
        • et al.
        Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56.
        Med Phys. 1997; 24: 1557-1598
        • Kari Tanderup
        • Hellebust Taran P
        • Lang Stefan
        • et al.
        Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer.
        Radiotherapy and Oncology. 2008; 89: 156-163
        • Paulsen Hellebust T
        • Kirisits Christian
        • Berger Daniel
        • et al.
        Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy.
        Radiotherapy and Oncology. 2010; 96: 153-160
        • Daniel Berger
        • Dimopoulos Johannes
        • Pötter Richard
        • et al.
        Direct reconstruction of the Vienna applicator on MR images.
        Radiotherapy and Oncology. 2009; 93: 347-351
        • Irene Humer
        • Kirisits Christian
        • Berger Daniel
        • et al.
        Improved source path localisation in ring applicators and the clinical impact for gynecological brachytherapy.
        Journal of contemporary brachytherapy. 2015; 7: 239-243
        • Jon Hansen
        • Dunkerley David
        • Bradley Kristin
        • et al.
        Comparison of catheter reconstruction techniques for the lunar ovoid channels of the VeneziaTM applicator.
        Journal of Contemporary Brachytherapy. 2020; 12: 383-392
        • Su Fan-chi
        • Li Xing
        • Huang Jessica Y.
        • et al.
        Dosimetric variations in gynecological HDR brachytherapy planning using a hybrid HDR applicator.
        International Journal of Radiation Oncology Biology Physics. 2020; 108: 484-485
        • Awunor O A
        • Dixon B
        • Walker C
        Direct reconstruction and associated uncertainties of 192Ir source dwell positions in ring applicators using gafchromic film in the treatment planning of HDR brachytherapy cervix patients.
        Physics in Medicine & Biology. 2013; 58: 3207-3225
        • Paulsen Hellebust Taran
        • Tanderup Kari
        • Bergstrand Eva Stabell
        • et al.
        Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation.
        Physics in Medicine & Biology. 2007; 52: 4893-4904
        • Holt JG
        AAPM Report No. 41: Remote afterloading technology.
        Med Phys. 1993; 20: 1761
        • Fuentemilla Naiara
        • Fernandez Aitor
        • Pellejero Santiago
        • et al.
        PO-0202 commissioning of a Venezia applicator: discrepancies between expected and actual source positions.
        Radiotherapy and Oncology. 2021; 158: 161-162
        • Li Xing
        • Su Fan-chi
        • Huang Jessica Y.
        • et al.
        Commissioning of Venezia lunar ovoid applicator using electron autoradiograph.
        Med Phys. 2020; 47: 722-723
        • Natasa Milickovic
        • Saeed Butt
        Commissioning of the Venezia cervix hybrid applicator.
        Jahrestagung der Deutschen Gesellschaft für medizinische Physik. 2019; : 18-21
        • Marc Morcos
        • Sheikh Khadija
        • Rezaee Mohammad
        • et al.
        Commissioning and loading pattern determination for Elekta's Venezia gynecological applicator for patients with advanced cervical cancer.
        Brachytherapy. 2018; 17: 36
        • Awunor O A
        • Berger D
        • Kirisits C
        A multicenter study to quantify systematic variations and associated uncertainties in source positioning with commonly used HDR afterloaders and ring applicators for the treatment of cervical carcinomas.
        Med Phys. 2015; 42: 4472-4483
        • Kathrin Kirchheiner
        • Nout Remi A
        • Lindegaard Jacob C
        • et al.
        Dose–effect relationship and risk factors for vaginal stenosis after definitive radio (chemo) therapy with image-guided brachytherapy for locally advanced cervical cancer in the EMBRACE study.
        Radiotherapy and oncology. 2016; 118: 160-166
        • Sandy Mohamed
        • Lindegaard Jacob C
        • Leeuw Astrid AC de
        • et al.
        Vaginal dose de-escalation in image guided adaptive brachytherapy for locally advanced cervical cancer.
        Radiotherapy and Oncology. 2016; 120: 480-485
        • Niroomand-Rad Azam
        • Chiu-Tsao Sou-Tung
        • Grams Michael P
        • et al.
        Report of AAPM task group 235 radiochromic film dosimetry: an update to TG-55.
        Med Phys. 2020; 47: 5986-6025
        • Niroomand-Rad Azam
        • Blackwell C R
        • Coursey B M
        • et al.
        Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine.
        Med Phys. 1998; 25: 2093-2115
      1. VeneziaTM Advanced Gynecological Applicator (instructions for use): appendix C- Measured source path for ring applicators and Advanced Gynecological Applicator.
        Elekta (Stockholm, Sweden). 2019;
        • Nazanin Hoshyar
        • Cohen Gilad
        • Voros Laszlo
        • et al.
        Commissioning of tandem and split ring applicator.
        Brachytherapy. 2016; 15: 153