Metal artifact reduction in cervix brachytherapy with titanium applicators using dual‐energy CT through virtual monoenergetic images and an iterative algorithm: A phantom study

  • Abolfazl Kanani
    Affiliations
    Ionizing and Non-Ionizing Radiation Protection Research Center (INIRPRC), School of Paramedical Sciences, Shiraz University of Medical Sciences, Shiraz, Iran
    Search for articles by this author
  • Mehran Yazdi
    Affiliations
    Signal and Image Processing Lab. (SIPL), School of Electrical and Computer Eng., Shiraz University, Shiraz, Iran
    Search for articles by this author
  • Amir M. Owrangi
    Affiliations
    Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, TX
    Search for articles by this author
  • Sareh Karbasi
    Affiliations
    Physics Unit, Department of Radio-oncology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
    Search for articles by this author
  • Mohammad Amin Mosleh-Shirazi
    Correspondence
    Corresponding author: Radio-oncology Department, Namazi Teaching Hospital, Shiraz University of Medical Sciences, Shiraz, 71936-13311, Iran. Tel: +98 71 36125316; fax: +98 71 36474320.
    Affiliations
    Ionizing and Non-Ionizing Radiation Protection Research Center (INIRPRC), School of Paramedical Sciences, Shiraz University of Medical Sciences, Shiraz, Iran

    Physics Unit, Department of Radio-oncology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
    Search for articles by this author

      ABSTRACT

      PURPOSE

      To evaluate an iterative metal-artifact reduction (iMAR) algorithm, dual-energy CT (DECT) through virtual monoenergetic images (VMI), and a combination of iMAR and DECT for reducing metal artifact severity (AS) induced by Fletcher titanium applicators used in cervix brachytherapy, the efficacy of which are hitherto unreported.

      METHODS AND MATERIALS

      120 kVp single-energy CT (SECT) (Siemens) of BEBIG tandem applicators, varying in shape (straight or curved) and diameter (3.5 mm or 5 mm) in a custom-made water-filled phantom, and their DECT images obtained from extrapolation of 80 kVp and 140 kVp, were reconstructed using four methods: DECT through VMI±iMAR, and SECT±iMAR. The DECT images were reconstructed monoenergetically at 70, 150, and 190 keV. AS was evaluated using measured values and statistical analysis.

      RESULTS

      iMAR, DECT, and combined DECT and iMAR reduced AS (p < 0.05). DECT had a lower AS than SECT, even without iMAR (p < 0.025). SECT+iMAR was more effective than DECT-iMAR with VMI at 70 and 190 keV (p < 0.05), whereas showing no statistically significant difference at 150 keV. With DECT and iMAR combined, AS was reduced more effectively compared to the SECT+iMAR or DECT alone. It also reduced the mean interobserver uncertainty by 0.2 mm.

      CONCLUSIONS

      These findings indicate that iMAR reduces the AS caused by Fletcher titanium applicators for both SECT and DECT, a combination of iMAR and DECT is superior to either strategy alone, and at low energies, DECT+iMAR also produces similar artifact reduction. These practical strategies promise more accurate source-position and structure definitions in CT-based gynecological brachytherapy treatment planning.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Brachytherapy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

      1. GLOBOCAN Cervical Cancer Fact Sheet. Available at http://globcan.iarc.fr/old/FactSheets/cancers/cervix-new.asp. Accessed: 25 March 2022.

        • Kim Y.J.
        • Kang H.C.
        • Kim Y.S
        Impact of intracavitary brachytherapy technique (2D versus 3D) on outcomes of cervical cancer: a systematic review and meta-analysis.
        Strahlentherapie und Onkologie. 2020; 196: 973-982
        • Narayan K.
        • Barkati M.
        • Van Dyk S.
        • et al.
        Image-guided brachytherapy for cervix cancer: from Manchester to Melbourne.
        Expert review of anticancer therapy. 2010; 10: 41-46
        • Haie-Meder C.
        • Potter R.
        • Van Limbergen E.
        • et al.
        Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group☆(I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV.
        Radiotherapy and oncology. 2005; 74: 235-245
        • Choong E.S.
        • Bownes P.
        • Musunuru H.B.
        • et al.
        Hybrid (CT/MRI based) vs. MRI only based image-guided brachytherapy in cervical cancer: Dosimetry comparisons and clinical outcome.
        Brachytherapy. 2016; 15: 40-48
        • Barrett J.F.
        • Keat N.
        Artifacts in CT: recognition and avoidance.
        Radiographics. 2004; 24: 1679-1691
        • Brown D.W.
        • Damato A.L.
        • Sutlief S.
        • et al.
        A consensus-based, process commissioning template for high-dose-rate gynecologic treatments.
        Brachytherapy. 2016; 15: 570-577
        • Kanani A.
        • Owrangi A.M.
        • Mosleh-Shirazi M.A.
        Comprehensive methodology for commissioning modern 3D-image-based treatment planning systems for high dose rate gynaecological brachytherapy: A review.
        Physica Medica. 2020; 77: 21-29
        • Xia D.
        • Roeske J.C.
        • Yu L.
        • et al.
        A hybrid approach to reducing computed tomography metal artifacts in intracavitary brachytherapy.
        Brachytherapy. 2005; 4: 18-23
        • Roeske J.C.
        • Lund C.
        • Pelizzari C.A.
        • et al.
        Reduction of computed tomography metal artifacts due to the Fletcher-Suit applicator in gynecology patients receiving intracavitary brachytherapy.
        Brachytherapy. 2003; 2: 207-214
        • Yazdi M.
        • Lari M.A.
        • Bernier G.
        • et al.
        An opposite view data replacement approach for reducing artifacts due to metallic dental objects.
        Medical physics. 2011; 38: 2275-2281
        • Axente M.
        • Paidi A.
        • Von Eyben R.
        • et al.
        Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy.
        Medical physics. 2015; 42: 1170-1183
        • Huang J.Y.
        • Kerns J.R.
        • Nute J.L.
        • et al.
        An evaluation of three commercially available metal artifact reduction methods for CT imaging.
        Physics in Medicine & Biology. 2015; 60: 1047
        • Subhas N.
        • Primak A.N.
        • Obuchowski N.A.
        • et al.
        Iterative metal artifact reduction: evaluation and optimization of technique.
        Skeletal radiology. 2014; 43: 1729-1735
        • Meyer E.
        • Raupach R.
        • Lell M.
        • et al.
        Normalized metal artifact reduction (NMAR) in computed tomography.
        Medical physics. 2010; 37: 5482-5493
        • Meyer E.
        • Raupach R.
        • Lell M.
        • et al.
        Frequency split metal artifact reduction (FSMAR) in computed tomography.
        Medical physics. 2012; 39: 1904-1916
        • van Elmpt W.
        • Landry G.
        • Das M.
        • et al.
        Dual energy CT in radiotherapy: current applications and future outlook.
        Radiotherapy and Oncology. 2016; 119: 137-144
        • Tahmasebi Birgani M.J.
        • Mahdavi M.
        • Zabihzadeh M.
        • et al.
        Simultaneous characterization of electron density and effective atomic number for radiotherapy planning using stoichiometric calibration method and dual energy algorithms.
        Australasian Physical & Engineering Sciences in Medicine. 2018; 41: 601-619
        • Kruis M.F.
        Improving radiation physics, tumor visualisation, and treatment quantification in radiotherapy with spectral or dual-energy CT.
        Journal of applied clinical medical physics. 2022; 23: e13468
        • Rousselle A.
        • Amelot A.
        • Thariat J.
        • et al.
        Metallic implants and CT artefacts in the CTV area: Where are we in 2020?.
        Cancer/Radiothérapie. 2020; 24: 658-666
        • Wang Y.
        • Qian B.
        • Li B.
        • et al.
        Metal artifacts reduction using monochromatic images from spectral CT: evaluation of pedicle screws in patients with scoliosis.
        European journal of radiology. 2013; 82: e360-e366
        • Bamberg F.
        • Dierks A.
        • Nikolaou K.
        • et al.
        Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.
        European radiology. 2011; 21: 1424-1429
        • Mallinson P.I.
        • Coupal T.M.
        • McLaughlin P.D.
        • et al.
        Dual-energy CT for the musculoskeletal system.
        Radiology. 2016; 281: 690-707
        • Bär E.
        • Schwahofer A.
        • Kuchenbecker S.
        • et al.
        Improving radiotherapy planning in patients with metallic implants using the iterative metal artifact reduction (iMAR) algorithm.
        Biomedical Physics & Engineering Express. 2015; 1025206
        • Kidoh M.
        • Nakaura T.
        • Nakamura S.
        • et al.
        Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR).
        Clinical radiology. 2014; 69: e11-e16
        • Elzibak A.H.
        • Kager P.M.
        • Soliman A.
        • et al.
        Quantitative CT assessment of a novel direction-modulated brachytherapy tandem applicator.
        Brachytherapy. 2018; 17: 465-475
        • Kim Y.J.
        • Cha J.G.
        • Kim H.
        • et al.
        Dual-energy and iterative metal artifact reduction for reducing artifacts due to metallic hardware: a loosening hip phantom study.
        American Journal of Roentgenology. 2019; 212: 1106-1111
        • Sun X.
        • Zhao Q.
        • Sun P.
        • et al.
        Metal artifact reduction using mono-energy images combined with metal artifact reduction software in spectral computed tomography: a study on phantoms.
        Quantitative Imaging in Medicine and Surgery. 2020; 10: 1515
        • Lim P.
        • Barber J.
        • Sykes J.
        Evaluation of dual energy CT and iterative metal artefact reduction (iMAR) for artefact reduction in radiation therapy.
        Australasian physical & engineering sciences in medicine. 2019; 42: 1025-1032
        • Pawałowski B.
        • Panek R.
        • Szweda H.
        • et al.
        Combination of dual-energy computed tomography and iterative metal artefact reduction to increase general quality of imaging for radiotherapy patients with high dense materials. Phantom study.
        Physica Medica. 2020; 77: 92-99
        • Beaulieu L.
        • Carlsson Tedgren A.
        • Carrier J.F.
        • et al.
        Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation.
        Medical physics. 2012; 39: 6208-6236
        • Sina S.
        • Faghihi R.
        • Meigooni A.S.
        • et al.
        Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs-137 brachytherapy source.
        Journal of Applied Clinical Medical Physics. 2011; 12: 183-193
        • Hakvoort E.T.
        • Wellenberg R.H.
        • Streekstra G.J.
        Quantifying near metal visibility using dual energy computed tomography and iterative metal artifact reduction in a fracture phantom.
        Physica Medica. 2020; 69: 9-18
        • Winklhofer S.
        • Hinzpeter R.
        • Stocker D.
        • et al.
        Combining monoenergetic extrapolations from dual-energy CT with iterative reconstructions: reduction of coil and clip artifacts from intracranial aneurysm therapy.
        Neuroradiology. 2018; 60: 281-291
        • Bongers M.N.
        • Schabel C.
        • Thomas C.
        • et al.
        Comparison and combination of dual-energy-and iterative-based metal artefact reduction on hip prosthesis and dental implants.
        PLoS One. 2015; 10e0143584
        • Lee K.Y.G.
        • Cheng H.M.
        • Chu C.Y.
        • et al.
        Metal artifact reduction by monoenergetic extrapolation of dual-energy CT in patients with metallic implants.
        Journal of Orthopaedic Surgery. 2019; 272309499019851176
        • Mangold S.
        • Gatidis S.
        • Luz O.
        • et al.
        Single-source dual-energy computed tomography: use of monoenergetic extrapolation for a reduction of metal artifacts.
        Investigative radiology. 2014; 49: 788-793
        • Wellenberg R.H.
        • Boomsma M.F.
        • van Osch J.A.
        • et al.
        Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses.
        European journal of radiology. 2017; 88: 61-70
        • Hokamp N.G.
        • Neuhaus V.
        • Abdullayev N.
        • et al.
        Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms.
        Skeletal radiology. 2018; 47: 195-201
        • May M.S.
        • Bruegel J.
        • Brand M.
        • et al.
        Computed tomography of the head and neck region for tumor staging—comparison of dual-source, dual-energy and low-kilovolt, single-energy acquisitions.
        Investigative radiology. 2017; 52: 522-528
        • Wu A.
        • Tang D.
        • Wu A.
        • et al.
        Comparison of the Dosimetric Influence of Applicator Displacement on 2D and 3D Brachytherapy for Cervical Cancer Treatment.
        Technology in Cancer Research & Treatment. 2021; 2015330338211041201
        • Tanderup K.
        • Hellebust T.P.
        • Lang S.
        • et al.
        Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer.
        Radiotherapy and Oncology. 2008; 89: 156-163
        • Kirisits C.
        • Rivard M.J.
        • Baltas D.
        • et al.
        Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.
        Radiotherapy and oncology. 2014; 110: 199-212