Impact of detector selection on commissioning of Leipzig surface applicators with improving immobilization in high-dose-rate brachytherapy

      Abstract

      PURPOSE

      Commission and treatment setup of Leipzig surface applicators, because of the steep dose gradient and lack of robust immobilization, is challenging. We aim to improve commissioning reliability by investigating the impact of detector choice on percentage depth dose (PDD) verifications, and to enhance accuracy and reproducibility in calibration/treatment setup through a simple and novel immobilization device.

      METHODS AND MATERIALS

      PDD distributions were measured with radiochromic films, optically stimulated luminescent dosimeters (OSLDs), a diode detector, and both cylindrical and parallel plate ionization chambers. The films were aligned to the applicators in parallel and transverse orientations. PDD data from a benchmarking Monte Carlo (MC) study were compared with the measured results, where surface doses were acquired from extrapolation. To improve setup accuracy and reproducibility, a custom-designed immobilization prototype device was made with cost-effective materials using a 3D printer.

      RESULTS

      The measured PDD data with different detectors had an overall good agreement ( < ± 10 % ). The parallel plate ionization chamber reported unreliable doses for the smallest applicator. There was no remarkable dose difference between the two film setups. The two-in-one prototype device provided a rigid immobilization and a flexible positioning of the applicator. It enhanced accuracy and reproducibility in calibration and treatment setup.

      CONCLUSION

      We recommend using radiochromic films in the transverse orientation for a reliable and efficient PDD verification. The applicator's clinical applicability has been limited by a lack of robust immobilization. We expect this economical, easy-to-use prototype device can promote the use of Leipzig applicators in surface brachytherapy.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Brachytherapy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Shah Chirag
        • Ouhib Zoubir
        • Kamrava Mitchell
        • Koyfman Shlomo A.
        • Campbell Shauna R.
        • Bhatnagar Ajay
        • Canavan Joycelin
        • et al.
        The American Brachytherapy society consensus statement for skin brachytherapy.
        Brachytherapy. 2020; 19: 415-426
        • Guinot Jose L.
        • Rembielak Agata
        • Perez-Calatayud Jose
        • Rodríguez-Villalba Silvia
        • Skowronek Janusz
        • Tagliaferri Luca
        • Guix Benjamin
        • et al.
        GEC-ESTRO ACROP recommendations in skin brachytherapy.
        Radiotherapy and Oncology. 2018; 126: 377-385
        • Ouhib Zoubir
        • Kasper Michael
        • Calatayud Jose Perez
        • Rodriguez Silvia
        • Bhatnagar Ajay
        • Pai Sujatha
        • Strasswimmer John
        Aspects of dosimetry and clinical practice of skin brachytherapy: The American Brachytherapy Society working group report.
        Brachytherapy. 2015; 14: 840-858
        • Likhacheva Anna O.
        • Devlin Phillip M.
        • Shirvani Shervin M.
        • Barker Christopher A.
        • Beron Phillip
        • Bhatnagar Ajay
        • Doggett Stephen W.
        • et al.
        Skin surface brachytherapy: A survey of contemporary practice patterns.
        Brachytherapy. 2017; 16: 223-229
        • Fulkerson Regina K.
        • Perez-Calatayud Jose
        • Ballester Facundo
        • Buzurovic Ivan
        • Kim Yongbok
        • Niatsetski Yury
        • Ouhib Zoubir
        • et al.
        Surface brachytherapy: Joint report of the AAPM and the GEC-ESTRO Task Group No. 253.
        Medical Physics. 2020; 47: e951-e987
        • Draeger Emily
        • Pinkham Daniel W.
        • Chen Zhe Jay
        • Tien Christopher J.
        Clinically-implementable template plans for multidwell treatments using Leipzig-style applicators in 192Ir surface brachytherapy.
        Brachytherapy. 2021; 20: 401-409
        • Tien Christopher J.
        • Pinkham Daniel W.
        • Chen Zhe Jay
        Feasibility of using multiple-dwell positions in 192Ir Leipzig-style brachytherapy surface applicators to expand target coverage and clinical application.
        Brachytherapy. 2020; 19: 532-543
        • Rivard Mark J.
        • Coursey Bert M.
        • DeWerd Larry A.
        • Hanson William F.
        • Saiful Huq M.
        • Ibbott Geoffrey S.
        • Mitch Michael G.
        • Nath Ravinder
        • Williamson Jeffrey F.
        Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.
        Medical physics. 2004; 31: 633-674
        • Evans M.D.
        • Yassa Mariam
        • Podgorsak Ervin B.
        • Roman Ted N.
        • Schreiner L.John
        • Souhami Luis
        Surface applicators for high dose rate brachytherapy in AIDS-related Kaposi's sarcoma.
        International journal of radiation oncology, biology, physics. 1997; 39: 769-774
        • Hwang Ming
        • Lin Shen-Yeh
        • Lin Li-Ching
        • Chuang Keh-Shih
        • Ding Hueisch-Jy
        Alternative effective modality of Leipzig applicator with an electron beam for the treatment of superficial malignancies.
        Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2003; 508: 460-466
        • Niu Hongquan
        • Hsi Wen C.
        • Chu James CH
        • Kirk Michael C.
        • Kouwenhoven Erik
        Dosimetric characteristics of the Leipzig surface applicators used in the high dose rate brachy radiotherapy: Dosimetric characteristics of Leipzig applicator in HDR brachytherapy.
        Medical physics. 2004; 31: 3372-3377
        • Pérez-Calatayud José
        • Granero Domingo
        • Ballester Facundo
        • Puchades Vicente
        • Casal Emilio
        • Soriano Angela
        • Crispín Vicente
        A dosimetric study of Leipzig applicators.
        International Journal of Radiation Oncology* Biology* Physics. 2005; 62: 579-584
        • Jeong Jeho
        • Barker Christopher A.
        • Zaider Marco
        • Cohen Gil'ad N.
        Impact of source position on high-dose-rate skin surface applicator dosimetry.
        Brachytherapy. 2016; 15: 650-660
        • Fulkerson Regina K.
        • Micka John A.
        • DeWerd Larry A.
        Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part II. High dose rate 192Ir sources.
        Medical Physics. 2014; 41022104
        • Sayler Elaine
        • Eldredge-Hindy Harriet
        • Dinome Jessie
        • Lockamy Virginia
        • Harrison Amy S.
        Clinical implementation and failure mode and effects analysis of HDR skin brachytherapy using Valencia and Leipzig surface applicators.
        Brachytherapy. 2015; 14: 293-299
        • Murphy Liam.
        The dosimetric effect of variations in source position on treatments using Leipzig-style brachytherapy skin applicators.
        Biomedical Physics & Engineering Express. 2020; 6015031
        • Niroomand-Rad Azam
        • Blackwell Charles Robert
        • Coursey Bert M.
        • Gall Kenneth P.
        • Galvin James M.
        • McLaughlin William L.
        • Meigooni Ali S.
        • Nath Ravinder
        • Rodgers James E.
        • Soares Christopher G.
        Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55.
        Medical physics. 1998; 25: 2093-2115
        • Niroomand-Rad Azam
        • Chiu-Tsao Sou-Tung
        • Grams Michael P.
        • Lewis David F.
        • Soares Christopher G.
        • Battum Leo J.Van
        • Das Indra J.
        • et al.
        Report of AAPM task group 235 radiochromic film dosimetry: an update to TG-55.
        Medical physics. 2020; 47: 5986-6025
        • Hermida-López M.
        • Lüdemann L.
        • Flühs A.
        • Brualla L.
        Influence of the phantom material on the absorbed-dose energy dependence of the EBT3 radiochromic film for photons in the energy range 3 keV–18 MeV.
        Medical Physics. 2014; 41112103
        • Tien Christopher Jason
        • Ebeling III, Robert
        • Hiatt Jessica R.
        • Curran Bruce
        • Sternick Edward
        Optically stimulated luminescent dosimetry for high dose rate brachytherapy.
        Frontiers in oncology. 2012; 2: 91
        • Jursinic Paul A.
        Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements.
        Medical physics. 2007; 34: 4594-4604
        • Sharma Renu
        • Jursinic Paul A.
        In vivo measurements for high dose rate brachytherapy with optically stimulated luminescent dosimeters.
        Medical physics. 2013; 40071730
        • Kry Stephen F.
        • Alvarez Paola
        • Cygler Joanna E.
        • DeWerd Larry A.
        • Howell Rebecca M.
        • Meeks Sanford
        • O'Daniel Jennifer
        • et al.
        AAPM TG 191: Clinical use of luminescent dosimeters: TLDs and OSLDs.
        Medical physics. 2020; 47: e19-e51
        • Roos M.
        • Derikum K.
        • Lange B.
        Eine neue Flachkammerkonstruktion fur die Elektronendosimetrie Mediz.
        Physik 1993: 24 Wissenschaftliche Tagung der Deutschen Geselleschaft fur Medicinische Physik. 1993;
        • Gersh Jacob A.
        • Best Ryan CM
        • Watts Ronald J.
        The clinical impact of detector choice for beam scanning.
        Journal of applied clinical medical physics. 2014; 15: 174-193
        • Casar Božidar
        • Gershkevitsh Eduard
        • Mendez Ignasi
        • Jurković Slaven
        • Saiful Huq M.
        Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations—perpendicular and parallel.
        Medical Physics. 2020; 47: 242-259
        • DeWerd Larry A.
        • Venselaar Jack LM
        • Ibbott Geoffrey S.
        • Meigooni Ali S.
        • Stump Kurt E.
        • Thomadsen Bruce R.
        • Rivard Mark J.
        Overview on the dosimetric uncertainty analysis for photon-emitting brachytherapy sources, in the light of the AAPM Task Group No 138 and GEC-ESTRO report.
        Metrologia. 2012; 49: S253
        • Casey Kevin E.
        • Alvarez Paola
        • Kry Stephen F.
        • Howell Rebecca M.
        • Lawyer Ann
        • Followill David
        Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry.
        Medical physics. 2013; 40112102
        • Palmer A.L.
        • Bradley D.A.
        • Nisbet A.
        Dosimetric audit in brachytherapy.
        The British journal of radiology. 2014; 8720140105
        • Villalba Silvia Rodríguez
        • Perez-Calatayud Maria Jose
        • Bautista Juan Antonio
        • Carmona Vicente
        • Celada Francisco
        • Tormo Alejandro
        • García-Martinez Teresa
        • et al.
        Novel simple templates for reproducible positioning of skin applicators in brachytherapy.
        Journal of Contemporary Brachytherapy. 2016; 8: 344