Dynamic intraoperative dosimetry for prostate brachytherapy using a nonisocentric C-arm
Abstract
Purpose
To evaluate a prototypical system of dynamic intraoperative dosimetry for prostate brachytherapy using registered ultrasound and fluoroscopy (RUF) with a nonisocentric C-arm (GE OEC, Salt Lake City, UT) and to compare intraoperative dosimetry of RUF as well as ultrasound-based seed localization (USD) with Day 0 CT dosimetry.
Methods
Seed positions were independently determined using RUF and USD. RUF uses a radio-opaque fiducial for registration to ultrasound and 3-dimensional reconstruction of seeds relative to prostate using nonisocentric C-arm fluoroscopy. Postimplant CT was performed on Day 0. Squared differences between dosimetric measures for RUF vs. CT and USD vs. CT were calculated and mean squared differences evaluated. Paired t test was used to evaluate which method was more closely aligned with CT. Accuracies of USD and RUF compared with CT were estimated using a nonparametric approach.
Results
Six patients were treated and compared with USD. RUF identified areas of underdosage intraoperatively in all patients and median 5 additional seeds were placed. In 40 of 42 measures, RUF was equally or more closely correlated with CT than USD. USD showed statistically significant variation from CT for 6 of 7 parameters compared with 1 of 7 parameters for RUF. Mean squared differences from CT were significantly smaller for RUF in 4 of 7 parameters compared with USD.
Conclusions
Dynamic intraoperative dosimetry is possible with a conventional nonisocentric C-arm. Compared with an USD method, RUF-based intraoperative dosimetry was more closely aligned with immediate postimplant CT. RUF identified areas of underdosage, which were not detected using USD.
Keywords: Prostate brachytherapy, Dynamic dosimetry, C-arm, Real-time dosimetry, Intraoperative
To access this article, please choose from the options below
This study was supported by Department of Defense W81XWH-05-1-0407 and W81XWH-06-1-0003, National Institutes of Health NIH 2 R44 CA099374-02/04, and National Science Foundation NSF EEC-9731478.
Conflicts of interest notification: Dr. E. Clif Burdette is employed by Acoustic Medsystems, which holds patents related to the registered ultrasound and fluoroscopy technology and Dr. Gabor Fichtinger holds patents related to the registered ultrasound and fluoroscopy technology. No other conflicts of interest exist.
PII: S1538-4721(10)00249-7
doi:10.1016/j.brachy.2010.06.005
© 2011 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.